Schlumberger Stabilock SI 4031
for the full repair see
youtu.be/AJSm5VRBqQc
Do these repairs at your own risk, this video is for information only.
RF Attenuator fault and Fix
-20dB chip found to be faulty, could not find an equivalent part so simply made a T pad with 2 x 39R and a shunt resistor of 10R, exact values should be 2 x 40.91R and 10.10R =-20dB.
Next job is to change around 70 x 10uf SMD caps!!
Fault
Test of the RF Attenuator 226-031S
TEST FREQUENCY =100MHz, LEVEL= 0dBA
Relays Energized/engaged by pushing the plunger on the end of the relay can. No power supplied.
INPUT OUTPUT
BU49 TO BU46 = -70dBA (N0 RELAYS ENERGIZED, Should be -50dBA) ++++
ENERGIZED OUTPUT BU46
RLS2 AND RLS3 = 0dBA (PATH STRAIGHT THROUGHT)
RLS2 = -30dBA (PATH THROUGH B3 30dBA)
RLS3 = -40dBA (PATH THROUGH B2 20dBA) ++++++
RLS2, RLS3, RLS4 = -30dBA (PATH THROUGH B4 30dBA)
RLS2, RLS3, RLS5 = -20dBA (PATH THROUGH B5 20dBA)
Above implies RLS2 (B2) which is reading -40dBA should be -20dBA +++++
Schlumberger Stabilock SI 4031
for the full repair see
youtu.be/AJSm5VRBqQc
Do these repairs at your own risk, this video is for information only.
RF Attenuator fault and Fix
-20dB chip found to be faulty, could not find an equivalent part so simply made a T pad with 2 x 39R and a shunt resistor of 10R, exact values should be 2 x 40.91R and 10.10R =-20dB.
Next job is to change around 70 x 10uf SMD caps!!
Fault
Test of the RF Attenuator 226-031S
TEST FREQUENCY =100MHz, LEVEL= 0dBA
Relays Energized/engaged by pushing the plunger on the end of the relay can. No power supplied.
INPUT OUTPUT
BU49 TO BU46 = -70dBA (N0 RELAYS ENERGIZED, Should be -50dBA) ++++
ENERGIZED OUTPUT BU46
RLS2 AND RLS3 = 0dBA (PATH STRAIGHT THROUGHT)
RLS2 = -30dBA (PATH THROUGH B3 30dBA)
RLS3 = -40dBA (PATH THROUGH B2 20dBA) ++++++
RLS2, RLS3, RLS4 = -30dBA (PATH THROUGH B4 30dBA)
RLS2, RLS3, RLS5 = -20dBA (PATH THROUGH B5 20dBA)
Above implies RLS2 (B2) which is reading -40dBA should be -20dBA +++++