We repair a defective HP8485A 26.5GHz RF Power Sensor that refuses to Zero with its meter HP437B. We also calibrate the N port cal signal against the Marconi 6204B Microwave Test Set, (Scalar Analyser) utilising a Marconi 6910 power sensor for reference since this Marconi kit is calibrated we use the Marconi test gear to calibrate the HP Meter & confirm its sensors are accurate. We setup the sweep gen on the Marconi 6204B Microwave Test Set & increase the resoloution of measurment to seceral decimal places so that accurate adjustment of the dbm level output of the 50MHz port is acumplished using settling after each micro adjustment of the level setting postentiomenter, this takes time vs temprature so isnt something to be rushed for absoloute accuracy.
We repair a defective HP8485A 26.5GHz RF Power Sensor that refuses to Zero with its meter HP437B. We also calibrate the N port cal signal against the Marconi 6204B Microwave Test Set, (Scalar Analyser) utilising a Marconi 6910 power sensor for reference since this Marconi kit is calibrated we use the Marconi test gear to calibrate the HP Meter & confirm its sensors are accurate. We setup the sweep gen on the Marconi 6204B Microwave Test Set & increase the resoloution of measurment to seceral decimal places so that accurate adjustment of the dbm level output of the 50MHz port is acumplished using settling after each micro adjustment of the level setting postentiomenter, this takes time vs temprature so isnt something to be rushed for absoloute accuracy.