Here we look at a Hp Agilent spectrum analyser 9khz to 3GHz model with a reported foul of not able to calibrate & RF level reading low. We embark on a path of verification, & a series of tests of internal circuits & modules inc the calibrator & 600MHz oscillator & reference oscillator, PLL circuits & 3rd converter analysis finally we diagnose the fault. During this process we use numerous test instruments to measure signals & help diagnose the fault.
Here we look at a Hp Agilent spectrum analyser 9khz to 3GHz model with a reported foul of not able to calibrate & RF level reading low. We embark on a path of verification, & a series of tests of internal circuits & modules inc the calibrator & 600MHz oscillator & reference oscillator, PLL circuits & 3rd converter analysis finally we diagnose the fault. During this process we use numerous test instruments to measure signals & help diagnose the fault.