Uploaded May 2021 | Updated November 2026, 4 days ago
In this episode Shahriar takes a look at a GW-Instek programmable AC/DC source with 1kVA, 200V capability. The instrument is ideal for power applications since it offers customized waveforms as well as AC+DC output. The unit shows COMMUNICATION ERROR 2 when powered which indicates an internal module-to-module fault. An X-Ray of the main processor board shows the various power supply planes and isolated sub-components communicating through optocouplers. The power supply design is reverse engineered and the fault is traced to a set of bad operational amplifiers. All broken components are replaced which restores the unit's functionality. The instrument is then tested to show some of its capabilities.
TheSignalPath.com
YouTube.com/TheSignalPath
Patreon.com/TheSignalPath
In this episode Shahriar takes a look at a GW-Instek programmable AC/DC source with 1kVA, 200V capability. The instrument is ideal for power applications since it offers customized waveforms as well as AC+DC output. The unit shows COMMUNICATION ERROR 2 when powered which indicates an internal module-to-module fault. An X-Ray of the main processor board shows the various power supply planes and isolated sub-components communicating through optocouplers. The power supply design is reverse engineered and the fault is traced to a set of bad operational amplifiers. All broken components are replaced which restores the unit's functionality. The instrument is then tested to show some of its capabilities.
TheSignalPath.com
YouTube.com/TheSignalPath
Patreon.com/TheSignalPath