Uploaded April 2026 | Updated November 2026, 6 days ago
Before you press start, here are some things to check on your AKTA (or other brand of FPLC).
First things first, ensure that the right buffer line is in the right buffer. But that’s not the only thing to look for buffer-wise. It’s critical that you don’t introduce any air into the system, so you want to ensure, right before the run:
- There are no air bubbles in your buffer lines
- You have plenty of buffer in your bottle for your run
- The buffer line is secured (either with the special bottle caps or tape)
- The bottom of the buffer line is right above the surface of the bottom of the bottle
- The buffer line isn’t curled up
Also, make sure you have enough empty tubes, etc. for collecting all of your fractions.
AKTA tutorials: youtube.com/playlist?list=PLUWsCDtjESrF66pHtzzddnKfSnYE7UcqD & thebumblingbiochemist.com/fplc-akta-practical-tutorials
lots more on protein purification: bit.ly/proteinpurificationtech & youtube.com/playlist?list=PLUWsCDtjESrEtqQkEXsQpTyRfMhywBVQ7
more practical lab tips & tricks: bit.ly/lab_tricks_page & youtube.com/playlist?list=PLUWsCDtjESrFEAWZCRKJL7sMc6a_KgfLU
more about all sorts of things: #365DaysOfScience All (with topics listed) 👉 bit.ly/2OllAB0 or search blog: http://thebumblingbiochemist.com
#scicomm #biochemistry #molecularbiology #biology #sciencelife #science #realtimechem
Before you press start, here are some things to check on your AKTA (or other brand of FPLC).
First things first, ensure that the right buffer line is in the right buffer. But that’s not the only thing to look for buffer-wise. It’s critical that you don’t introduce any air into the system, so you want to ensure, right before the run:
- There are no air bubbles in your buffer lines
- You have plenty of buffer in your bottle for your run
- The buffer line is secured (either with the special bottle caps or tape)
- The bottom of the buffer line is right above the surface of the bottom of the bottle
- The buffer line isn’t curled up
Also, make sure you have enough empty tubes, etc. for collecting all of your fractions.
AKTA tutorials: youtube.com/playlist?list=PLUWsCDtjESrF66pHtzzddnKfSnYE7UcqD & thebumblingbiochemist.com/fplc-akta-practical-tutorials
lots more on protein purification: bit.ly/proteinpurificationtech & youtube.com/playlist?list=PLUWsCDtjESrEtqQkEXsQpTyRfMhywBVQ7
more practical lab tips & tricks: bit.ly/lab_tricks_page & youtube.com/playlist?list=PLUWsCDtjESrFEAWZCRKJL7sMc6a_KgfLU
more about all sorts of things: #365DaysOfScience All (with topics listed) 👉 bit.ly/2OllAB0 or search blog: http://thebumblingbiochemist.com
#scicomm #biochemistry #molecularbiology #biology #sciencelife #science #realtimechem