Technical Service & Repair Report

Equipment Under Test (EUT): Marconi 2955A Radio Test Set

Calibration / Verification Standard: HP8920B Radio Test Set

Initial Status: Unit fails all internal self-tests upon arrival.
1. Initial Assessment & Visual Inspection

The initial diagnostic strategy was to isolate the failure to either the Signal Generator or the Transmitter (TX) Monitor subsystem. Upon removing the chassis covers, a visual inspection revealed extensive evidence of poor-quality legacy repair attempts, including:

An undocumented, improvised wiring loom soldered directly to the rear of the keyboard assembly, featuring an unanchored, free-floating microswitch.

Handwritten fault notes inscribed directly onto internal screening cans, alongside a business card taped inside detailing prior symptoms.

Widespread workmanship issues, notably multiple missing chassis and component screws critical to structural and shielding integrity.

2. Phase 1: Power Rail Rectification & VCO Realignment

Initial circuit tracing identified a power supply failure on the +12V rail feeding the Signal Generator amplifier circuitry.

Component Replacements: Diagnostics isolated a short-circuited electrolytic capacitor (220μF, 16V) which had subsequently caused a 33μH inductor to burn out. Both components were replaced, successfully restoring the +12V rail and the primary RF Signal Generator output.

VCO Troubleshooting: Following power restoration, a secondary fault emerged: the 200MHz Voltage-Controlled Oscillator (VCO) was completely inactive. Investigation revealed that a previous technician had misadjusted the VCO trimmer capacitor, pushing the circuit out of its lock range.

Resolution: The official factory realignment procedure was executed to re-net the VCO trimmer. This restored the 200MHz reference, enabling correct synthesis across the lower frequency bands (400kHz to 86MHz).

3. Phase 2: Frequency Counter Repair & Attenuator Overhaul

A subsequent execution of the internal self-test suite yielded a majority pass, though several error codes remained outstanding.

Frequency Counter Diagnostics: The error codes pointed to the Frequency Counter stage. Signal level checks were conducted on the OM345 wideband amplifier IC located on the Frequency Counter PCB. While the +12V supply rail was verified as stable, the IC exhibited a severe gain deficit. The OM345 was diagnosed as defective and replaced, which successfully cleared all related self-test errors.

Attenuator Restoration: Cross-conformance testing was conducted using an HP8920B Radio Test Set. Evaluation of the Marconi’s RF Signal Generator output revealed erratic behavior from the step attenuator. The attenuator assembly was removed and systematically stripped down. Two critical screws were found missing from the actuator body from a prior repair. The internal contacts were thoroughly cleaned, the unit was reassembled with replacement hardware, and testing confirmed full, stable attenuation linearity.

4. Phase 3: Front Panel & Keypad Remediation

To resolve mechanical entry faults, the front panel was disassembled to access the keypad PCB.

Modification Reversal: The non-standard wiring loom and microswitch modification were completely desoldered and removed to restore the board to OEM specifications.

Switch Refurbishment: The internal tactile switches for the RF GEN, TX, RX, LEVEL, and FREQ keys were disassembled, chemically cleaned of oxidation, reassembled, and verified for continuity.

5. Final Validation and Closeout

The front panel assembly and chassis covers were securely reinstalled using correct torque specifications and replacing all previously missing hardware.

A final comprehensive self-test sequence was initiated, resulting in a 100% Pass across all parameters. Bench verification against the HP8920B confirmed that all RF output and monitoring specifications met factory standards. Anti-tamper security labels were applied to the chassis seams, and the asset was formally signed off as fully operational and ready for service.
MARCONI 2955A Radio Test Set Repair, multiple faults, OM345, Step Attenuator, Keybaord, & VCO fault.Repair of Air Traffic Control Radio Park Air VHF Airband transceiver inc testingKenwood TS2000 Very Poorly radio attempted repair. Water Damage & faults everywhere.ICOM 736 Spec Test for owner before sale.SPECTRUM ANALYSER RHODE & SCHWARZ REPAIR FSEB30 7GHz model. Strip down analysisSaxophone repair & explanation of the keys & notes also a play test Conn Selmer prelude modelHP8920 Radio Test Set PSU repair PART 3 FINALROHDE & SCHWARZ CMS33 Radio Tester, with Avionics options Review, Strip-down, Analysis walk & talk.Spectrum Analyser demo Aeroflex 3251 3GHz plus repair 150Watt RF attenuator.HAMFEST RADIO RALLY UK Market Rasen Lincolnshire APRIL 2022 run by Lincoln Shortwave Radio Club.Marconi 6200a PART3 Microwave Radio Test Set repairCold War PYE Radio VHF Transceiver & Radiation Dosimeter from the 1970s from a Nuclear Bunker.