while testing an input signal i pressed the "Auto-Scale button" after a few clicks the scope locked up, and now will only show the below Error code. "SOFTWARE ERROR - PLEASE RECORD THESE NUMBERS 9A8E 00102008 00FF"
checked the power supply module voltages both in circuit and with a 2 ohm load. All voltages found to be correct.
Replaced the bios 3.0v battery and new install of the ROM and SYSTEM files from a 1.44 MByte floppy disk. Tried the two versions available, Unreleased software rev T2.26 and Released software rev 2.00.
Both installed OK, but i still get the above Software error code.
The mother board is 98% surface mount components, i have checked the caps for leakage and can see no signs of any burnt tracks, removed and cleaned the ADC hybrid modules. Part 2 is to test the ROMs!!!
internal PSU part No. 0950-2369
HP 54540C Oscilloscope, Fault: SOFTWARE ERROR. Part 1Elecraft KX3 fitting KXAT3 internal ATUiOptron MiniTower Altitude Locking Knob DIY UpgradeFluke 196C ScopeMeter, Basic Probe Cal & Functions.CREWORKS 8x14 Mini Lathe, Quick Change Tool Post Holder, QCTPSurecom SW-102HF, unbox and test, Marconi 2966ATUXING Pcp Air Compressor, TXET063, FAIL, 20 Bar out of calibrationBulldog ST Alpha Airsoft Gun, review & chronograph testIC-7700, FTM400, IC-271H using RODE NT1-A & Behringer MIC2200, FBQ1502HD MX822Amstrad 8060 Fix Power Button, Volume pot and Replace Cassette Belts, Part 3MotoRadds Motorcycle Chin Mount on Schuberth C4,  tested on BMW R1200GS with GoPro