Function Gen says 1V, Scope says 2V. Which one's right?

Click to subscribe! ► bit.ly/Scopes_Sub

The function generator is set to have a 1 V output, but the oscilloscope measures a 2 Volt signal. Why? When set to 50 ohm mode, the function generator actually outputs double the specified input voltage. It does this because it expects to see a 50 ohm load on the output, creating a 50/50 voltage divider circuit.

Questions or comments? Ask me on Twitter:
@DanielBogdanoff twitter.com/DanielBogdanoff

Learn more about using oscilloscopes:
oscilloscopelearningcenter.com

Check out the EEs Talk Tech podcast:
youtube.com/playlist?list=PLzHyxysSubUnAMeCIi2-S0Vm7YtSAGqx_

The digital storage oscilloscope we used: keysight.com/en/pcx-x205209/infiniivision-4000-x-series-oscilloscopes?cc=US&lc=eng

#oscilloscope #oscilloscopes #electronics #electricalengineering #impedance #impedancematching #50ohm #50ohmimpedance #highimpedance #outputimpedance #functiongenerator
Dont Fall Into This Output Impedance Trap!Big News From Keysight!How to make a St. Patricks Day waveform and Todays Scope Month Winners! (17-Mar-2017)Antistatic Rolex Mod (ESD Strap Build)Determining Oscilloscope Measurement Quality - Scopes University - (S1E4)AC coupling & Todays Scope Month Winners! (9-Mar-2017)Heres What Imaginary Power Looks LikeThe Electro Turbo EncabulatorPre-Compliance Conducted Emissions Test - The ABCs of EMC (E03)Narrow vs. Wide IF Bandwidth - How it Impacts Your VNA Measurements + Wave WinnersMeasuring Voltage with an Oscilloscope - The Keysight 2-Minute Guru (s2e5)Attenuators and Preamps - Oscilloscope Front End Design (part 3)

Don't Fall Into This Output Impedance Trap! @KeysightLabs

SHARE TO X SHARE TO REDDIT SHARE TO FACEBOOK WALLPAPER