Extinction coefficients can be given in terms of molarity if you know the sequence of the protein. but when you don’t (and even sometimes when you do) they’re given in mg/mL terms. They tell you the absorbance value (A) that corresponds to 10 mg/mL (1%) or 1 mg/mL (0.1%). Those percentages come from the weight/volume percentage convention that a 1% solution corresponds to 1 g/100 mL - more on why here: bit.ly/37LsTrq

More on protein concentration measuring: bit.ly/proteinmeasuring & youtu.be/nSkdz_Mso7E

You can calculate the estimated extinction coefficient using free online software tools like Expasy ProtParam. I say estimated because context matters - the local environment around the absorbing part can influence how eager it is to absorb a photon

When I do this for BSA I see this:

Extinction coefficients:
Extinction coefficients are in units of M-1 cm-1, at 280 nm measured in water.
Ext. coefficient 42925
Abs 0.1% (=1 g/l) 0.638, assuming all pairs of Cys residues form cystines
Ext. coefficient 40800
Abs 0.1% (=1 g/l) 0.607, assuming all Cys residues are reduced

First it tells me the values under oxidizing conditions and below that it tells me the values under reducing conditions (the intracellular environment is reducing and we usually add reducing agents like DTT or β-mercaptoenthanol) to protein solutions to keep them happy outside the cell). It tells me this because cysteine crosslinks can also absorb, where applicable.

Then I can plug this into Beer’s law (or have the computer do it for me) if I measure the absorbance.

You measure this absorbance using something called a spectrophotometer. Basically it shines light through a solution and measures to what extent different wavelengths make it through (are transmitted) versus don’t make it through (are absorbed). We looked at how this can be converted into concentration of solute (dissolved molecules) using Beer’s Law bit.ly/2N4nzXE

The equation is: A = εcl

A = absorbance

ε = extinction coefficient (aka molar absorptivity coefficient) - specific for particular molecule & particular wavelength; units of L mol-1cm-1

c = concentration (in mol/L) - this is molarity - a mole is just a chemist’s “baker’s dozen” - it’s Avogadro’s number (6.022 x 10^23) of something - solute molecules or donuts, it’s just a number bit.ly/2r4RnrX

l = path length (in cm)

For Beer’s Law, you only need the absorbance at a single wavelength - but there’s much more to learn if you look at the whole spectrum (or at least a couple key values). Because molecules have overlapping spectra (e.g. both DNA & RNA absorb light of 260nm wavelength) you look to ratios.

As I discussed, proteins absorb most strongly at 280, and this is where we typically calculate from. Proteins also absorb at 230nm and that absorbance is from the generic backbone part - corresponds to absorbance by the peptide bonds linking the letters. These peptide bonds also have resonance, but not as much as rings do, and they absorb ~190-230nm.

Because DNA absorbs so strongly at UV260, where protein doesn’t, it’s relatively easy to see if you have DNA in your protein prep, but it’s harder to tell if you have protein in your DNA prep - 260 will dominate the 260/280 ratio

Moral of the story: there’s no “one right way” to count your proteins & it’s important to carefully choose the one you use!

For more practical protein-purification posts (and background/theory), check out this page on my blog where I’ve collected some of my protein purification posts. bit.ly/proteinpurificationtech

More on protein concentration measuring: bit.ly/proteinmeasuring & youtu.be/nSkdz_Mso7E

More on light absorption & Beer’s law: bit.ly/dnauvbeer

More on the NanoDrop: bit.ly/nanodrop_science

More on ExPasy ProtParam: bit.ly/uniprotprotparam

more on topics mentioned (& others) #365DaysOfScience All (with topics listed) 👉 bit.ly/2OllAB0

#scicomm #biochemistry #molecularbiology #biology #sciencelife #science #realtimechem
Converting UV280 absorbance to protein concentration - finding conversion values & using on NanoDropPaper walk/talk-through on articles exploring use of the CRISPR-Cas12a2 system in eukaryotic cellsLab tip: Use a quick Bradford assay to tell when to move on in your protein column chromatographyThere are often many right ways to do the same overall lab experimentTips for setting up PAGE gel casting cassettes – without leaksProtein crystal sizeKey information to know about amino acids & tips for rememberingBiochemistry Word Parts: a non-exhaustive list of key prefixes, suffixes, roots, etc. you may seeA peek at strategic SEC peak pickingNew outline & resource guide, this one an overview of recombinant protein techniquesNegative staining in the context of cryo-EMInterpreting Table 1 crystallography data
the bumbling biochemist |

Converting UV280 absorbance to protein concentration - finding conversion values & using on NanoDrop

SHARE TO X SHARE TO REDDIT SHARE TO FACEBOOK WALLPAPER