This unit did not have any output and intermittently showed a couple of different errors during self-test. The problem was traced to a batch of faulty 301 ohm resistors. 10 out of 48 failed. After replacing the faulty resistors the instrument seems to work fine.
Correction: there are 50 resistors of this type on the board (not 48 as I said in the video)
#138 - Keysight 33250A function generator repair#35 - M-Audio Axiom Pro 49 repair#58 - Fluke 97 ScopeMeter battery upgrade to Li-ion#62 - Shure ULX transmitter repair#53 - Precision thermometer for data logging#40 - EDC 520A DC calibrator repair and adjustment#144 - Shure BLX receiver repair fail#72 - Fluke A90 current shunt#112 - Korg ER-1 mkII repair#28 - Tektronix PS2520 lab power supply repair and calibration#149 - PreSonus NSB 16.8 repair#68 - Using Agilent 661x for testing Li-ion batteries